CAUTION: Luminex calibrators and verifiers contain sodium azide. Sodium
azide is toxic and may also react with lead and copper plumbing
to form highly explosive metal azides. On disposal, flush drains
with a generous amount of cold water to prevent azide build-up.
Consult the manual guideline Safety Management No. CDC-22,
Decontamination of Laboratory Sink Drains to remove Azide
Salts (Centers for Disease Control, Atlanta, Georgia, April 30,
1976).
Luminex
®
FLEXMAP 3D
®
Hardware
The Luminex
®
FLEXMAP 3D
®
system includes the following hardware:
The FLEXMAP 3D
®
instrument
Personal computer (PC) and accessories
Stereo speakers
Power cables
Three sample probes
Off-plate reagent area
96 and 384 heater blocks
Empty cubitainer for waste
Sheath fluid intake line
Waste fluid outtake line
USB communication cable
Barcode reader (optional)
Sample probe height alignment kit and 384-well plate Probe Height Adjustment tool
Monitor arm
NOTE: For information on calibrating the probe correctly, refer to the
FLEXMAP 3D 384-Well Plate Probe Height Adjustment Tool section.
Recommended Additional Equipment
Successful operation of the
Luminex
®
FLEXMAP 3D
®
system may require additional
equipment.
Uninterruptible Power Supply (UPS) or Surge Protector
Luminex recommends using either an uninterruptible power supply (UPS) or a surge
protector to protect your system from power outages. Use a UPS that provides 1300 watts for
at least 45 minutes. Select a surge protector that fits your requirements with regard to
electrical environment, endurance, suppressed voltage rating, and method of protection. The
surge protector requires three outlets and a minimum rating of 1500 watts. Either piece of
equipment should be listed by Underwriters Laboratory (UL) or a similar listing body, certified
by Canadian Standards Association (CSA), and marked Conformité Europeénne (CE) for
nondomestic use.
For research use only. Not for use in diagnostic procedures.